13.2 Options of the Patient Menu and Corresponding
Descriptions
Option
Range
Off, 1 ~ 5
Humidifier
Off, 1 ~ 3
Reslex
Ramp
0 ~ Max
Time
Ramp
Delay
On / Off
2000-01-01
Date
—
2099-12-31
00:00
Time
—
23:59
Brightness
High / Low
Full Face;
Mask
Nasal;
Type
Pillow;
Other
iCode,
iCode
iCode QR,
iCode QR +
0 ~
Use Time
50000 h
About
——
E-20C-H-O / E-20CJ-H-O User Manual V2.3
There are five humidity levels available. As the numbering
increases, the humidity rises accordingly. "Off" means the
humidifier is turned off
This feature enables the device to automatically reduce the
treatment pressure when the patient exhales, so as to make
the user more comfortable. The higher the numbering is, the
more pressure the device reduces. "Off" means this feature
is disabled
In order to increase comfort and help the patient fall asleep
easily, the pressure can increase gradually, when the Ramp
feature is enabled. The ramp time during which the initial
pressure rises to the prescribed treatment pressure can be
adjusted. As you turn the Knob
the numbering increases or decreases by five minutes. The
screen displays a real-time countdown of the remaining
ramp time in minutes
When the humidifier is on, this feature allows the airflow to
continue for about 15 minutes at a low pressure (about 2 hPa)
after you press the Knob
will blow off the vapor left in the humidifier to avoid any
damage to the device. When this feature is set to "Off," which
means it is disabled, the airflow stops instantly after you press
the Knob
Setting date by adjusting this option
Setting time by adjusting this option
Setting screen brightness by adjusting this option
There are three mask types available, namely Full Face
(full-face mask), Nasal (nasal mask), and Pillow (nasal pillow
mask). patient can choose other suitable masks as well.
When selecting masks other than the above three types of
BMC masks, the patient can identify the masks as other
iCode provides access to the patient's compliance data
during a recent time period. The iCode QR mode displays
data in sequences of characters, and the iCode QR + mode
displays data in two-dimensional codes
Use Time displays how long has the device been used by the
user. The use time can be erased
Displays related information of the device (Model, SN,
Version, ID). This is read-only and cannot be edited
21 / 35
Description
to the nearest point,
to discontinue treatment. This