When the humidifier is used, the
humidified air tends to condense
Water in mask
in the cold tube and mask if the
room temperature is low
Nasal, sinus, or
Sinus or middle ear inflammation
ear pain
The
uncomfortable
Discomfort due to
treatment pressure is higher than
inability to adapt
13 hPa. However, the treatment
to the treatment
pressure is determined according
pressure
to the patient's conditions, and
cannot treat sleep apnea if the
treatment pressure is set too low
Probably because the patient
sleeps with his or her mouth
Obstructive sleep
open, and the pressurized air
apnea symptoms
goes out via the mouth, leading
recur
to blockage in the respiratory
tract
The device is too
The
noisy
properly
Air delivered from
The air inlet of the device may be
the
device
is
partially
abnormally hot
insufficient airflow into the device
E-20C-H-O / E-20CJ-H-O User Manual V2.3
patient
will
when
tube
is
not
connected
blocked,
leading
28 / 35
Turn
the
humidity
down, or raise the room
temperature.
Place the tube under the quilt,
or use the tube cover.
Hang the tube loosely, and the
lowest part of the tube should
be lower than the patient's
head
Contact
your
immediately
feel
It takes a maximum of four
the
weeks to adapt to pressurized
air.
Relax and breathe through the
nose. If the problem still
exists, contact your physician
Use a chin strap to prevent the
mouth from opening during
sleep, or use a full-face mask.
Contact your physician for
details
Reconnect the tube properly
Replace the air filter (see 15.6
Replacing the Air Filter), and
clean the air inlet
Place the device in an area
to
where air flows freely, and
make sure the device is at
least 20 centimeters away
from the wall, curtain, or other
things
setting
physician