APPENDIX 2
Guidance and manufacture's declaration – electromagnetic emissions-for all EQUIPMENT and SYS-
TEMS
Guidance and manufacture's declaration – electromagnetic emission
The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
customer of the user of the Pulse Oximeter should assure that it is used in such and environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Guidance and manufacture's declaration – electromagnetic immunity –for all EQUIPMENT and SYS-
TEMS
Guidance and manufacture's declaration – electromagnetic immunity
The Pulse Oximeter is intended for use in the electromagnetic environment specified below. The cus-
tomer or the user of Pulse Oximeter should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Power frequency
(50Hz)
magnetic field
IEC 61000-4-8
Compliance
Electromagnetic environment – guidance
Group 1
The Pulse Oximeter uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not like-
ly to cause any interference in nearby electronic equipment.
Class B
The Pulse Oximeter is suitable for use in all establishments,
including domestic establishments and those directly connect-
ed to the public low-voltage power supply network that sup-
plies buildings used for domestic purposes.
IEC 60601
Compliance
test level
level
8 kV contact
8 kV contact
15 kV air
15 kV air
30 A/m
30 A/m
ENGLISH
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floor are covered with synthetic material, the
relative humidity should be at least 30%. the
manufacturer may recommend the ESD precau-
tionary procedures to user.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typ-
ical commercial or hospital environment.