Alarms and troubleshooting
Model
Alarm(s)
Data connection lost at the
E970
beginning of the calibration.
Connection lost during the
E971-
process.
E975
Calibration device internal error.
Lina MB
Model selector in the wrong
E981
position.
Lina
Bad serial connection.
PRO13-
Fault/leak in the pneumatic
003
E982-
circuit.
E985
Electronic boards and/or
MS
sensors.
E986-
Calibration unsuccessful
E987
A31
Calibration unsuccessful
Lisa MB
Model selector in the wrong
A90
position
Power failure or data connection
A91
lost
Cause(s)
Turn off/on the calibration device.
Check the data cable connection.
Turn off/on the calibration device.
Check the data cable connection.
Pre-existing malfunctioning: check the "CYCLE HISTORY" menu of the sterilizer.
Turn off/on the calibration device.
Check the model selector is in the correct position.
Turn off/on the calibration device.
Check the sterilizer circuit and the test connection for leaks. After fixing the leak, run a
vacuum test.
Pre-existing malfunctioning: check the "CYCLE HISTORY" menu of the sterilizer.
in Service mode select MENU/CONFIG and confirm the following settings:
PROCESS SENSOR: PT1
LIQUID SENSOR: NOT INSTALLED
AIR DETECTOR: NOT INSTALLED
Turn off/on the calibration device
Check the sterilizer circuit and the test connection for leaks. After fixing the leak, run a
vacuum test.
Turn off/on the calibration device
Check the model selector is in the correct position
Turn off/on the calibration device
Check the data cable connection