4 In order to align the pyranometer horizontally, rotate the leveling feet until at least half the spirit level
bubble is in the inner ring.
4 Fix the pyranometer with the screws, making sure that the pyranometer retains the correct alignment.
4 To prevent corrosion between the screws and the pyranometer housing, make sure that the nylon
insulating rings are fixed.
4 Insert the connector with the cable into the pyranometer's connection socket.
4 Tighten the locking ring hand tight.
NOTE! The seal may be damaged by overtightening!
4 Fix the cable in such a way that the cable doesn't move or cast a shadow on the pyranometer.
4 Run the cable downwards so that water can flow off.
4 Fix the sun shade.
SMP12 Smart Pyranometer | EN
Assembly | 5