Safety And Precautionary Measures - MSA Altair 5X Manual De Instrucciones

Ocultar thumbs Ver también para Altair 5X:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Never block pump inlet, except to perform a sampling
system safety test.
Have a trained and qualified person interpret
instrument readings.
Risk of Explosion: Do not remove battery pack,
recharge Li Ion battery, or replace alkaline batteries in a
hazardous location.
Do not alter or modify instrument.
Use only MSA-approved sampling lines.
Do not use silicone tubing or sampling lines.
Wait sufficient time for the reading; response times vary
based on gas and length of sampling line.
Do not use the instrument for prolonged periods in an
atmosphere containing a concentration of fuel or solvent
vapors that may be greater than 10% LEL.
INCORRECT USE CAN CAUSE DEATH OR SERIOUS
PERSONAL INJURY.
This Class A digital apparatus complies with Canadian ICES-003.

1.2 Safety and Precautionary Measures

Carefully review the following safety limitations and
precautions before placing this instrument in service.
Incorrect use can cause death or serious personal injury.
Check function (see section 3.6) each day before use. MSA
recommends carrying out a routine inspection prior to each day's
use.
Perform a Bump Test (see section 3.7) before each day's use to
verify proper instrument operation. The instrument must pass the
bump test. If it fails the test, perform a calibration (see section 3.8)
before using the instrument.
The ALTAIR 5X MultiGas Detectors are designed to detect gases
and vapors in air only.
Perform a Bump Test more frequently if unit is subjected to
physical shock or high levels of contaminants. Also, perform a
Bump Test more frequently if the tested atmosphere contains the
following materials, which may desensitize the combustible gas
1-2
" WARNING

Publicidad

Tabla de contenido
loading