Hinweise zur
elektromagnetischen
Recommended separation distances between portable and mobile RF communications equipment and the
The VITOM
®
3D exoscope model TH 200 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output power
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters [m] can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts [W] according to the
transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
and people.
Information on
electromagnetic
Table 6
®
VITOM
3D exoscope model TH 200
according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
D
D
0.12
0.38
1.2
3.8
12
electromagnética (CEM)
[m]
800 MHz to 2.5 GHz
F
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23