14
EMISSION TEST
RF emissions
CISPR 11/EN 550011
RF emissions
CISPR 11/EN 550011
Harmonics emissions
IEC61000-3-2/EN
61000-3-2
Voltage fluctuation/
flicker emissions
IEC 61000-3-2/
EN61000-3-3
Guidance and manufacture's declaration - electromagnetic immunity
The bath lift RS900 is intended for use in the electromagnetic environment specified below. The
customer or the user of the above mentioned product should assure that it is used in such envi-
ronment.
IEC/EN 60601
Immunity test
TEST LEVEL
Electrostatic
+/-6KV contact
discharge(ESD)
+/-8KV air
IEC 61000-4-2
+/-2KV
Electrical fast
for power supply lines
transient/burst
+/-1KV
IEC61000-4-4
for input/output lines
+/-1KV for L-N
Surge IEC
+/-2KV
61000-4-5
for lines to earth
<5% UT(>95% dip in
UT)
for 0.5 cycle
Voltage dip and
40%UT (60%dip in
interruption
UT)
IEC 61000-4-11
for 5 cycles
<5% UT (>95%dip in
UT)
for 5 cycles
Power fre-
quency
3 A/m
magnetic field
IEC61000-4-8
Note
- UT is the AC mains voltage prior to application of the test level.
COMPLIANCE
Group 1
The product use RF Energy only for its internal
Class B
function. Therefore ,its RF emissions are very low
and not likely to cause any interference in nearby
electronic equipment. The RS900 is suitable for use
Class A
in all establishments and those directly connected
to the public low-voltage power supply network that
supplies building used for domestic purpose
Complies
Compliance level
+/-6KV
contact
+/-2KV for
power
supply lines
+/-1KV for L-N
<5% UT
(>95% dip in UT)
for 0.5 cycle
<5% UT
(>95%dip in UT)
for 5 cycles
N/A Not contain
component
sensitive to
magnetic field
ELECTROMAGNETIC
ENVIRONMENT- GUIDANCE
Electromagnetic
environment - guidance
The RS900 is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-
voltage power supply network that
supplies building used for domestic
purpose