Specifications - Chauvin Arnoux C.A 1864 Manual De Instrucciones

Termómetros infrarrojo
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

English
Français
appears in the lower left corner of the display, press the
UP and down buttons to adjust the desired values.
4.5
ERROR MESSAGES
HOLD
4
Exceeding of the maximum
value measurable
The maximum measurable
value is 1000° C / 1832 ° F,
which if exceeded causes the
instrument to display the
screen shown opposite.
5.1

SPECIFICATIONS

Measurement range : -50 to +1000° C (-58 to +1832° F)
Resolution : 0,1° C/0,1° F
Polarity indication : automatic (minus only)
Accuracy :
-50 to –20° C (-58 to –4° F):
-20 to 200° C (-4 to 392 ° F):
200 to 538° C (392 to 1000 ° F):
538 to 1000 ° C(1000 to 1832 ° F): ± 3.5% of the re ading ± 5 ° C(±9 ° F)
Field of view :
Response time t
Spectral reponse : 8 to 14µm
Emissivity : adjustable from 0,1 to 1
Screen lighting : adjustable by the user
ε = 0,9
° C
OL
AVG
5 SPECIFICATIONS
30/1 (C.A 1864)
: < 1s
90
HOLD
ε = 0,9
- OL
3
AVG
Exceeding of the minimum
value measurable
The minimum measurable
value is -50° C / - 58 ° F, which
if exceeded causes the
instrument to display the
screen shown opposite.
±5° C (±9° F)
± 1.5% of the readin g + 2 ° C (±3.6 ° F);
± 2.0% of the read ing +2 ° C(±3.6 ° F);
50/1 (C.A 1866)
C.A 1864 – C.A 1866
° C
24

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

C.a 1866

Tabla de contenido