Table specifications
Dimension
Details
Width
887 mm (35 inches) (not including wheel position)
Depth
600 mm (24 inches) (not including wheel position)
Table height range
725 mm to 1205 mm (29 inches to 47 inches)
Monitor Support Arm radius
1.
514 mm (20 inches)
Monitor mount type
VESA MIS-D, 75, C (hole pattern: 75 mm x 75 mm)
Monitor Support Arm load
3 kg (6.6 lbs) to 5 kg (11 lbs)
Table weight
53 kg (116.8 lbs)
Maximum load on table
65 kg (143.3 lbs)
Total mass
118 kg (260.1 lbs)
The table provides wheelchair accessibility.
Environmental considerations (ambient)
Do
NOT
operate the table in an environment that exceeds the environmental limits.
Relative Humidity
Status
Temperature
(Non-condensing)
Normal use
+10°C to +40°C
75% (maximum)
1. The radius is the maximum distance that the Monitor Support Arm extends beyond the width of the table.
Guidance and manufacturer declaration – electromagnetic
immunity
IEC60601 compliance
Immunity test
Electromagnetic environment-guidance
test level
Electrostatic
Floors should be wood, concrete or ceramic tile. If
±6 kV contact
discharge
floors are covered with synthetic material, the
±8 kV air
relative humidity should be at least 30%.
IEC61000-4-2
±2 kV for power supply
Electrical fast
lines
Mains power quality should be that of a typical
transient/burst
commercial or hospital environment.
±1 kV for input / output
IEC61000-4-4
lines
Surge
±1 kV line to line
Mains power quality should be that of a typical
commercial or hospital environment.
IEC61000-4-5
±2 kV line to earth
<5% U
T
(>95%dip in U
)
T
for 0.5 cycle
Voltage dips, short
40% U
interruptions and
T
(60% dip in U
)
voltage variations
T
for 5 cycles
Mains power quality should be that of a typical
on power supply
70% U
commercial or hospital environment.
lines
T
(30% dip in U
)
T
IEC61000-4-11
for 25 cycles
<5% U
T
(>95% dip in U
)
T
for 5 seconds
Power frequency
Power frequency magnetic fields should be at levels
(50/60Hz)
3 A/m
characteristic of a typical location in a typical
Magnetic field
commercial or hospital environment.
IEC61000-4-8
Note: U T is the AC mains voltage prior to the application of the test level.
Guidance and manufacturer declarations - electromagnetic
emissions
IEC60601
Emissions test
compliance
test level
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A
IEC61000-3-2
Voltage fluctuations/
Pst
flicker emissions
Dmax
IEC61000-3-3
Operations
The following procedures provide more information about using your table.
Atmospheric
Pressure
Adjusting the table height
70kPa to 106kPa
Ensure the cables are
WARNING
Select
to move the table up.
l
Select
to move the table down.
l
Moving the table
The table has lockable wheels to prevent it moving during use.
1. Remove all equipment from the table. Refer to the specific equipment instructions for details.
2. Lower the height of the table to its lowest position for optimal stability and unobstructed view.
3. Unplug the table mains cable from the wall socket.
4. Place all cables on the table. Ensure there are no trailing cables.
5. Pull the wheel lever up to unlock the wheel. Repeat for each wheel.
6. Carefully move the table to its new location.
7. When in position, push the wheel lever down to lock the wheel. Repeat for each wheel.
8. Ensure the table works as expected.
General cleaning
Be careful
NOT
CAUTION
The table does not require cleaning before each use.
In a medical environment, if any part of the table or equipment becomes dirty or soiled with bodily fluids,
follow the general cleaning procedures and your internal clinic or hospital decontamination procedures.
Electromagnetic environment-guidance
The Optos Table uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are
NOT
cause any interference in nearby electronic equipment.
The Optos Table is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
NOT
pulled while the table height is being adjusted.
to damage the table or equipment when cleaning.
Adjust the arm tension
To adjust the Monitor Support Arm tension, with a monitor fitted, follow these steps:
1. Bend the main arm downwards to expose the bolt access cover.
2. Remove and retain the access cover.
likely to
3. Using an Hex key (4mm) turn the bolt as required.
a. Turn the bolt clockwise to increase the tension and prevent the arm dropping.
b. Turn the bolt anti-clockwise to decrease the tension and prevent the arm rising.
4. Refit the access cover.
Adjust the monitor mount plate
To adjust the monitor mount plate tension follow these steps:
1. Using an Hex key (4mm) turn the tension bolt either way as required:
a. Turn "+" to increase tension.
b. Turn "-" to decrease tension.