The SpO
Kit is ready to use immediately when you connect it to the main device via the
2
Communications Port.
The model of the SpO
2
OXYGEN SATURATION.
Attach its sensor to the patient's index finger or any other finger.
The sampling rate of the SpO
The value of SpO
and PR is calculated by the average of the former eight pulse waveforms.
2
If the SpO
Kit is in an abnormal state, the value of SpO
2
The screen of the main device then displays the Main Interface shown in Fig. 11-14 (only
applies to E-20A-H-O), or the Main Interface shown in Fig. 11-15 (only applies to E-20AJ-H-O).
The patient's blood oxygen saturation and pulse rate can be clearly seen during the course of
therapy.
WARNINGS!
• Change the measurement point regularly according to the patient's conditions after prolonged
use. Change the measurement point, check the patient's skin integrity and circulatory
conditions, and make the right adjustments at least every three hours.
• Excessive ambient light, excessive motion, use of intravascular dyes, poorly perfused finger,
extreme finger sizes or improper placement may degrade the SpO
the accuracy of the measurement.
• Nail polish or false nails should be removed before the finger sensor is used, or it may cause
erroneous measurements results.
• Overly low blood pressure, overly low systolic blood pressure, severe anemia, or hypothermia
may cause erroneous measurements results.
• The SpO
Kit is designed for use with this device only.
2
• Verify the compatibility of the device and SpO
the patient.
E-20A-H-O / E-20AJ-H-O User Manual V2.3
Kit is KS-CM01. The SpO
signal is about 50 Hz, and the update rate of the frame is 1 Hz.
2
Fig. 11-14
Fig. 11-15
Kit before use, otherwise it may cause injury to
2
16 / 36
Kit is calibrated to display FUNCTIONAL
2
will be blank.
2
Kit's performance or affect
2