Installation; First Installation; Safety Precautions During The Installation - Mectron Piezosurgery touch Manual De Uso

Ocultar thumbs Ver también para Piezosurgery touch:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 142
04 Û INSTALLATION
04.1 Û FIRST INSTALLATION
The device must be installed in a comfortable
place suitable for its use.
The PIEZOSURGERY® touch can be purchased
ready for it to be used instantly or may have to
be enabled by typing an activation key.
04.2 Û SAFETY PRECAUTIONS DURING INSTALLATION
WARNING: Contraindications.
Interference with other equipment.
Though compliant with the standard IEC
60601-1-2, the PIEZOSURGERY® touch may
nonetheless interfere with other devices
nearby.
The PIEZOSURGERY® touch must not
be used near to or stacked on other
devices. However, if this were to prove
necessary, you must check and monitor
correct operation of the device in that
configuration.
WARNING: Contraindications.
Interference from other equipment. An
electrical scalpel or other electro-surgical
units near the PIEZOSURGERY® touch
device may interfere with its correct
operation.
CAUTION: The electrical system of the
premises in which the device is installed
and used must be compliant to the norms
in force and to the relative electrical safety
precautions.
CAUTION: To avoid any risk of electric
shock this device must be grounded.
WARNING: Do not install the device in
places where there is a risk of explosion.
The device cannot operate in environments
where anesthetic or flammable mixtures
are present.
04 Û INSTALLATION
In the case that your device requires an
activation key, the procedures to follow may
vary from country to country.
Always refer to your retailer to obtain
information on the above.
CAUTION: Always position the device
in way so that the power switch is easily
reachable, since this switch is considered as
a load-break switch.
WARNING: When the lid of the
peristaltic pump is open, the foot pedal
of the PIEZOSURGERY® touch must not
be operated. The parts in movement may
cause harm to the operator.
WARNING: Install the device in a place
protected against collisions or against
accidental sprays of water or liquids.
WARNING: Do not install the device
above or near heat sources. Foresee
adequate air circulation around the device
when installing it. Leave adequate space,
especially near the fan placed on the back
part of the device.
CAUTION: Do not expose the device to
direct sunlight or to sources of UV light.
CAUTION: The device can be
transported, but it must be handled with
care when it is displaced. Position the foot
pedal on ground, so that it can only be
activated intentionally by the operator.
CAUTION: Before connecting the
handpiece cord to the device, make sure
that the electrical contacts are perfectly
dry. If need be, dry them with compressed
air.
EN
11

Publicidad

Tabla de contenido
loading

Tabla de contenido