04 Û INSTALLATION
- position the tube in the pump;
6
- close the pump cover completely;
WARNING: When the lid of the
peristaltic pump is open, the foot pedal
of the PIEZOSURGERY® touch must not
be operated. The parts in movement may
cause harm to the operator.
Connect the silicon tube end to the flow
7
adjuster;
Connect the flow adjuster to the liquid bag
8
used for the treatment (bag not supplied).
Hang the bag to its specific support rod;
Connect the foot pedal to the back of
9
the device in the socket marked with the
symbol
by means of the plug of the
pedal cable, until you hear a "click" sound;
PLEASE NOTE The foot pedal comes
equipped with a bracket that allows it to
be moved to the place most suitable for the
operation, without the need to use your
hands to move it;
The bracket can also be positioned
horizontally if it is not used;
EN
13