To fill the irrigation circuit, use the PUMP
function by selecting PUMP/CLEAN on the
touch keyboard: all the other selection
options present on the display are disabled,
and the PUMP/CLEAN writing flashes;
EN
While the PUMP/CLEAN wording is flashing,
press the foot pedal once and release it;
the PUMP/CLEAN stops flashing and the
irrigation circuit starts to fill up;
As soon as the peristaltic pump starts, the
entire scale of values of the "irrigation"
section lights up and during the liquid
passage, the value of the irrigation shifts
from 6 to 0;
The cycle can be interrupted as soon you see
liquid outflowing from the PIEZOSURGERY®
touch handpiece, by pressing PUMP/CLEAN
again or, alternatively, by pressing the foot
pedal. The PUMP function is disabled and
the keyboard is enabled again, and displays
the last setting used;
Screw the chosen insert onto the
PIEZOSURGERY® touch handpiece, until it
reaches the limit stop;