Stryker XPRT Manual De Uso página 73

Superficie de apoyo terapéutico. integración con fl27 (2155/2156)
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 293
XPRT™ THERAPEUTIC SUPPORT SURFACE (CONTINUED)
Recommended separation distances between portable and mobile
The XPRT™ is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled . The customer or the user of the XPRT™ can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the XPRT™ as recommended below, according to the maximum output power of the communications
equipmen
t .
Rated maximum output
power of transmitter
W
0 .01
0 . 1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter manufacturer .
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies .
NOTE 2
These guidelines may not apply in all situations . Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people .
www.stryker.com
EMC Information
RF communications equipment and the XPRT™.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d=1 .2
1 .2
0 .38
1 .2
3 .8
12
2950-109-005 REV B
m
80 MHz to 800 MHz
d=1 .2
0 .2
0 .38
1 .2
3 .8
12
800 MHz to 2 .5 GHz
d=2 .3
0 .23
0 . 7 3
2 .3
7 .3
23
Return To Table of Contents
1-69
English

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

2950

Tabla de contenido