PREVENTIVE MAINTENANCE
WEEKLY MAINTENANCE (CONTINUED)
REMOVING END SLEEVES
(Model FLU115D & FLU115DE)
NOTE:
The sleeve in these photos has been
modified (cut away) for clarity.
Remove the top sleeves to gain access to
the inside of the unit. Refer to
Pull the medium back and away from the
arm ports.
NOTE:
It is extremely important to move the medium away from the ports to
prevent the medium entering the space between the inner and outer
walls of the unit when the sleeves are removed and replaced.
Dust away all Cellex medium from inside
the sleeve.
Pull the retainer ring from the sleeve and
lay aside for re-installation.
page
26.
27 27
Fluidotherapy® - Dry Heat Therapy
Pull the sleeve away from the sleeve
extrusion in the inner housing of the unit.
NOTE:
There is a ring folded into the sleeve that
fits over the sleeve extrusion on the inside
of the unit.
From the outside of the unit, reach into the
sleeve and remove the other sleeve retainer
ring.
Cellex medium entering the heat
chamber of the unit can cause
severe injury to patients as well as
smoke damage to the facility and
the Fluidotherapy unit.
Carefully pull the sleeve and ring out of the
unit, avoiding Cellex medium from entering
the cavity between the two housings of the
unit.