PREVENTIVE MAINTENANCE
WEEKLY MAINTENANCE (CONTINUED)
Remove the ring from the folded end of the
sleeve.
Launder sleeves as described on
Vacuum out any of the Cellex medium that
may have entered the cavity between the
inner and outer walls of the unit.
Removal of all Cellex medium that has entered the cavity is essential to
maintain proper and safe operation of the unit. If the medium reaches
the heating chamber of the unit, immediately call a certified technician
for removal of any suspected medium in the heat chamber. Do not place
the unit into service if medium is suspected to be in the heat chamber.
page
26.
28
Fluidotherapy® - Dry Heat Therapy
INSTALLING 115 Model SLEEVE
Fold the elastic end of the sleeve over
the ring.
Place sleeve through port hole.
Place the ring/sleeve assembly over the
sleeve extrusion on the inside of the unit.
NOTE:
Position the sleeve seam toward the top of
the unit and push the ring/sleeve assembly
completely onto the sleeve extrusion.
Install the inner sleeve retainer ring making
certain the ring split will be toward the top of
the unit.
Push the retainer ring over the ring/sleeve
assembly until it is completely seated.
NOTE:
The retainer ring should fit tightly over the
sleeve. If necessary, use a pair of pliers to
tighten the ring in the area where it slides over
the sleeve.