Descargar Imprimir esta página

Dräger X-am 5000 Serie Instrucciones De Uso página 28

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 76
en
Intended use
Definitions of alert icons
The following alert icons are used in this document to provide and highlight areas
of the associated text that require a greater awareness by the user. A definition
of the meaning of each icon is as follows:
!
WARNING
Indicates a potentially hazardous situation which,
if not avoided, could result in death or serious injury.
!
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in
physical injury, or damage to the product or environment. It may also be used to
alert against unsafe practices.
Indicates additional information on how to use the product.
2
Intended use
Portable gas detection instrument for the continuous monitoring of the
concentration of several gases in the ambient air within the working area and in
explosion-hazard areas.
Independent measurement of up to 5 gases, in accordance with the installed
Dräger sensors.
3
Tests and Approvals
Marking
Copies of the name plate and the declaration of conformity are provided in the
enclosed supplementary documentation (order no. 90 33 890). Do not stick
anything on the name plate on the gas detector.
The technical suitability tests are valid for the X-am 5000 gas detector and the
calibration cradle. The explosion-protection approvals are only valid for the
X-am 5000 gas detector; the calibration cradle must not be used in the Ex zone.
The BVS 10 ATEX E 080 X technical suitability test is based on the adjustment
with the target gas.
28

Publicidad

loading

Este manual también es adecuado para:

X-am 5600