ProMinent DULCOTEST PER Serie Instrucciones De Servicio página 37

Sensor de peróxido de hidrógeno
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Conditions Operation of the sensor is stable (no possible drift or fluctuating
Zero point calibration A zero point calibration is only required at the lower limit of the
Slope test
ProMinent
®
measured values during a minimum period of 5 minutes). This is
generally guaranteed when the following conditions are fulfilled:
• the H
O
concentration of the sample water is isochronically
2
2
sufficiently constant (please note the sensor's response time of
8 min!)
• the relevant running-in time has been allowed (see section 8.1
Running-in time).
• permitted flow is present in the in-line probe housing (see section
14 Technical data).
• Temperature compensation is given between sensor and sample
water (wait approx. 15 minutes).
measuring range.
IMPORTANT
• After an initial commissioning, check calibration after 24
hours.
• Repeat calibration if the H
than 15 % from the reference value.
On photometry:
If not yet installed, install the sensor into the in-line probe DLG III
or DGM (see Chap. 6 Mounting)
Perform sampling. This must be performed in direct proximity to
the sensor. Recommendation: in case of the in-line probe DGM
use the sampling valve (see Fig. 2 and Order notes, Chap. 15).
Determine the H
2
Photometer (please observe operating instructions! E.g. use the
Photometer DULCOTEST
Set the determined H
device according to its operating instructions (see operating
instructions DULCOMETER
Chap. 8, Complete operating menu, Setting menu "Calibrate
H
O
").
2
2
O
concentration varies by more
2
2
O
concentration as quickly as possible with a
2
®
DT3).
O
concentration in ppm at the control
2
2
®
D1C, measured variable H
Commissioning
O
,
2
2
37

Publicidad

Tabla de contenido
loading

Tabla de contenido