Descargar Imprimir esta página

Ritter Concept Ultimate Comfort Smart Instrucciones De Uso página 59

Publicidad

Ultimate Comfort
Client (dental practice)
Device
Device type:
Dentale Behandlungseinheit
Manufacturer:
Ritter Concept GmbH
Type:
Serial No:
Netzspannung:
..........V
Safety inspection according: EN 62353  /   V DE 0751
Visual inspection
Protective earth conductor not demaged (for insulation class 1 only)
Cover and mechanical parts not demaged
Insulating parts not demaged
Device connecting cables including plugs without defects
Labels exist or completed, nameplate correctly
Others
Measurement
Protective earth resistance
Insulating resistance
Test voltage
Device leakage current (Equivalent device leakage current)
Encl. leakage current (Housing leakage current)
Leakage current of application part (Equivalent patient leakage current)
Test voltage leakage current (nominal Voltage ±10%)
Functional testing
Function and safety testing without defects
The unit has safety deficiencies:
Short-circuit L-N
Body circuit L/N-SL
Test cycle:......
Next test date:...................
Tester:
Place, Inspection date
© Ritter Concept GmbH
C20-0094-1801EN
Inspection protocol
for
safety inspection according DIN EN 62353
months
Signature
Sicherheitstechnische Kontrolle
Contractor
Used measurem ent device
Device type:
Electrical Safety Tester
Manufacturer:
Type:
Insulation class:
1
Type of applied parts:
Typ BF
measured value
<0,3 Ω
>2 MΩ
~500 V
<10 mA
<5 mA
<5 mA
min.:-10% U
max.+10% U
limit value
O.K./not O.K.
Netz
Netz
Anhang
Appendix 59

Publicidad