IMPORTANT: While the concentrator is powered from the DC power supply and operating in con-
tinuous mode at setting 2, the battery will not charge.
IMPORTANT: The DC power supply should be utilised on vehicles with the proper power output
rating.
IMPORTANT: The vehicle should be running while utilising the DC adapter to power the Zen-O
Portable Oxygen Concentrator.
To maximise battery life and run time, avoid letting the battery deplete and use while con-
EN
nected to a power source whenever possible. The internal battery will automatically charge
whenever the concentrator is connected to a power source. You can use the device while the
battery is charging. The LCD display will indicate whether the device is operating on battery
or external AC power.
The fully charged battery will retain some level of charge for up to thirty days in this device
when not in use - see Caution below for battery removal/storage recommendation.
IMPORTANT: Battery damage may result if the concentrator's battery is allowed to discharge com-
pletely.
IMPORTANT: After 300 charge/discharge cycles, the battery capacity will be at least 80% of its
original capacity. Replace the battery when the reduced battery life is affecting your mobility.
CAUTION: Remove battery if this device is not going to be used for more than seven days.
Store battery in a cool, dry place.
CAUTION: Check that this device operates on battery after disconnecting from the power
source.
IMPORTANT: When not using the battery inside the unit, be sure to store it in the protective sleeve
that was provided with the original package.

6.3. Nasal Cannula

Only use a nasal cannula with the following specifications:
• 2.1 m (7ft) or 7.6m (25ft) long
• High flow
• Crush resistant
• Large internal diameter bore
• Straight non-tapered tips
• Suitable for up to 15 litres per minute (lpm) at a max. pressure of 3.6 psi
• Meets substance compatibility of IEC/EN 60601-1
CAUTION: Only use approved accessories with this device. Refer to the approved accessories
guide for a complete list of accessories and cannula approved for use with this device. Us-
ing unapproved accessories or cannula may impair the performance of this device, including
flow rate or oxygen purity.