Alarm Indicators - GCE Zen-O RS-00500 Manual De Usuario

Ocultar thumbs Ver también para Zen-O RS-00500:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 150
The following general security screening considerations apply to Zen-O™ POCs:
1. The limit of one carry-on and one personal item (e.g., purse, briefcase, or computer case)
does not apply to medical supplies, equipment, mobility aids, and/or assistive devices
carried by and/or used by a person with a disability.
2. If a person has medical documentation regarding their medical condition or disability,
they can present this information to the screener to help inform him or her of the person's
situation. This documentation is not required and will not exempt a person from the secu-
rity screening process.
EN
9 .

Alarm Indicators

If the Zen-O™ Portable Oxygen Concentrator detects an alarm condition, it will indicate the
alarm visually and audibly within 10 seconds. There are four levels of alarms: critical high pri-
ority, high priority, medium priority, and low priority.
Each is indicated differently by the backlit display; yellow, and red LEDs; and buzzer, as in-
dicated below. In each case, the alarm message and power status will override the current
display.
IMPORTANT: All alarm conditions and parameters are factory preset; conditions and parameters
cannot be changed or adjusted by the user.
IMPORTANT: The alarm system is tested during the startup sequence. You should see all alarm
lights briefly turn on and the audible alarm indicator chirp.
Alarm status
Critical high
priority
High priority
Medium priority Three beeps per burst, burst
Low priority
IMPORTANT: If two alarm conditions exist at the same time, the highest priority alarm is indicated.
If two or more alarm conditions of equal priority exist at the same time, the most recent one will
be displayed.
IMPORTANT: The most recent alarms indicated by the device are logged for reference by service
personnel. This log is maintained even if the device is powered down or if power is lost for any
other reason.
Audible Tone
Ten beeps per burst, burst
repeats every 3 seconds.
Ten beeps per burst, burst
repeats every 3 seconds.
repeats every 8 seconds
Three beeps per burst, burst
repeats every 10 minutes
22/292
Visual Indicator
Solid red LED and device
shuts of automatically
Flashing red LED
Flashing yellow LED
Solid yellow LED
Mute Time
20 minutes
20 minutes
8 hours
24 hours

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido