9.1.5. Other Messages

Message
External power has been discon-
nected; unit is now running on
battery.
Displayed while unit goes through
its power-down sequence.
Displayed as the battery menu item
when there are no communications
with the battery.
Displayed percentage of battery
charge if at least 10% and there is
no external power connected.
NN% displays the current battery
charge level.
Displayed when battery charge
is greater than 10% but less than
100% and there is external power
connected.
Battery charge is less than 10% and
there is external power connected.
The patient's average breath rate
when the device is delivering the
maximum amount of oxygen and
the bolus is reduced. If no breaths
are detected, the most recent
breath rate is shown.
A previously set alarm has been
automatically cleared.
Description
25/292
Action
No action is required.
No action is required.
Verify that the battery pack is
correctly installed. Contact your
distributor if the battery is fully
inserted and the message contin-
ues to be displayed longer than 30
seconds.
Message is displayed when battery
button is pressed.
Message is displayed when battery
button is pressed.
Message is displayed when battery
button is pressed.
Reduced activity level. Be
sure air inlet/outlet has not been
blocked.
IMPORTANT: The message will
automatically clear when the device
returns to normal operation.
No action required.