TROUBLESHOOTING
8
TROUBLESHOOTING
1
Device does not turn on
Flow: Cannot remove hand piece
2
cartridge nut
Flow: Gutta-percha does not flow through
3
cannula
4
Pack: Hand piece does not work
5
Pack: The tip does not get warm
6
Flow: Delivery plunger does not retract
7
Flow: Motor stops turning
END-IFU-CALAMUS-DUAL-V06-WEB-EUR-Multilingual-06-2022
Problem
Solution
Verify that the power cord is plugged into
the device and into wall outlet.
Unplug device and check fuse. If fuse
is blown, replace it.
See "Replacing the
fuses".
Press Return button and allow drive
mechanism to back out in order to relieve
pressure on cartridge nut.
Verify that the cannula is not kinked -
avoid over bending of cannula.
Verify that the device has reached
operational temperature for gutta-percha
to flow.
Increase temperature if needed.
Make sure the tip is sitting in the hand
piece correctly.
Replace tip.
Check temperature settings.
Replace tip with a new one.
If, after pushing the Return button, the
gutta-percha indicator does not fully
retract to the end of the indicator window,
push the Return button again.
If the motor stops turning, the cannula
may have been placed too firmly in the
canal. Reduce the pressure applied to the
cannula and the motor may begin turning
again.