Operating Information
Alarm
Pen Home
(Down Arrow)
On/Off
On/Off
The ON/OFF key turns the unit on and off. Alarm
and calibration settings are restored to the most
recent values when the unit is turned on. If the
unit is using AC power and power fails, the unit
will switch to battery power without interrupting
the recording capabilities. If power fails when the
battery is not present, the pen will remain on the
chart in recording position.
Pen Home (also Down Arrow)
If the pen is located outside the chart, press the
HOME key to move pen to recording position. Press
the HOME key with the pen in recording position,
and the pen will move to the outside of the chart
and up the pen lift ramp (home).
The PEN HOME key is also used as a down arrow
key to decrease values when setting alarms or cali-
bration values.
Chart Rotation (also Up Arrow)
The CHART ROTATION key sets the chart to the
desired time setting. Press the CHART ROTATION
key to rotate chart. Continue pressing until the
chart reaches the desired time setting. (The red pen
aligned with the timing clip on the 1 thermocouple
unit and the blue pen aligned with the timing clip
on the 2 thermocouple unit.) The chart rotation will
reach full speed after a few seconds holding down
the key. Do NOT attempt to rotate the chart
manually. The CHART ROTATION key is also used
as an up arrow key to increase values when setting
alarms or calibration values.
Alarm (Optional)
The ALARM key turns the alarm on or off. When
the alarm is on the display will flash the MAX &
MIN indicators. Holding the ALARM key for 2
seconds will cause the unit to enter into the alarm
set mode. For a visual alarm the display will read
MIN if a minimum alarm is activated or MAX if the
maximum alarm is activated. The alarm condition
will also trigger an audio alarm if the audio alarm is
enabled (dip switch #8 is off). The alarm will sound
until it is disabled, or until the alarm condition no
longer applies.
LED
(Up Arrow)
Chart
Rotation
Figure 3
5
Alarm (2 thermocouples)
Holding the ALARM key for 2 seconds will cause
the unit to enter into the alarm set mode for Pen 1.
Pressing and holding the Alarm key a second time
will put the unit in Alarm Set mode for Pen2.
The PEN HOME and CHART ROTATION keys
are used for setting alarms. The PEN HOME
key decreases the alarm value and the CHART
ROTATION key increases the alarm value.
1.
To set the alarms press the ALARM key and
hold for 2 seconds. The alarm minimum set
point will be displayed with A and MIN will
also be displayed.
2.
Use the HOME key to decrease the minimum
alarm setting and the CHART ROTATION key
to increase the minimum alarm setting.
3.
Press ALARM key a second time and the alarm
maximum set point will be displayed along
with A and MAX indicating that you can now
set the maximum alarm. Use the same proce-
dure as in step #2 for setting the maximum
alarm setting. If your KT6 only has one pen
go to step 6.
4.
Press and hold the ALARM key for 2 sec-
onds and the Pen 2 alarm minimum set
point will be displayed along with A and
MIN indicating that you can now set the
minimum alarm. Use the same procedure
as in Step #2 for setting the minimum
alarm setting.
5.
Press alarm key a second time and Pen 2
maximum set point will be displayed along
with A and MAX indicating that you can
now set the maximum alarm. Use the
same procedure as in step #2 for setting
the maximum alarm setting.
6.
Press the ALARM key again to exit the
alarm setting mode.
7.
To activate the alarm, press the ALARM
key.
LED
The LED will be green whenever the unit is acti-
vated. The LED will flash green during initialization
and warm-up. The LED showing red indicates the
unit is in calibration mode. The LED turns orange
during shut down. The LED will also flash when the
unit is on if the AC power has failed at any time
since the unit has exited the warm-up period.