3.1.4 Inadmissible Operating Conditions
Operating the unit under improper conditions is not permitted as the operator's
safety cannot be guaranteed. Some operating conditions not permitted are the
following:
Using the unit for a purpose other than its intended use
Using of the unit when any part of it is damaged, not working properly, the
electrical installation is not correct, or the electrical insulation is damaged
Protective or safety equipment is non-functional, defective, improperly
installed, or missing
The unit or operating parameters were modified without consulting the
manufacturer
Operation in areas exposed to explosion hazards
Operation without a cooling media not recommended by the manufacturer
Any equipment connected to RS232 must be certified to IEC 61010-1/ IEC
60950-1/ IEC 623681 should be connected to the DB-9 port.
Only JULABO approved cable must be used to connect Flow meter to the
unit. Refer to Spare Parts to order this cable.
High Temperature Limit on Coolant Return - The maximum allowable
coolant return (inlet) is 50°C. It is the responsibility of the customer to
ensure that this limit is not exceeded. JULABO recommends a thermal
shutoff be used in conjunction with the equipment being connected to the
Recirculating Chiller. Methods include:
A method to turn off the heat load from injecting additional heat
o
into the coolant. This will allow the Recirculating Chiller to reduce
the temperature of the coolant further until stable.
A method that stops or restricts the coolant flow into the
o
Recirculating Chiller. This will result in a Coolant High Pressure
alarm, which will shut down the system for safety.
A method that interrupts the power to the Recirculating Chiller.
o
If any of these methods are activated, please reference the Alarms section.
NOTE
The manufacturer is not liable for damage occurring when using the unit in a
way it was not intended. This also voids JULABO' warranty.