Installation Procedure
8.1
Connect Hoses
The coolant hoses are connected to the unit by means of quick disconnect
couplings from Colder Products. Coolant inlet and coolant outlet are labeled
with respective symbols.
NOTE
Risk of damage by using improper or faulty coolant hoses
This may lead to damage to persons, damage to property, or corrosion
damage.
When choosing coolant hoses pay attention to burst pressure and
compatibility with coolant.
Only use coolant hoses without any signs of damage.
If water is being used as coolant, ensure that non-transparent
hoses are used to prevent algae growth in the water. Otherwise,
appropriate additives must be used
NOTE
When connecting the coolant hoses pay attention to flow direction. Follow
the documentation released by the manufacturer of the unit to be cooled.
Transparent coolant hoses increase algae growth and biofouling of the
components in the unit, and this reduces the performance of the unit. Thus,
only use non-transparent coolant hoses.
Installation Procedure
83