Descargar Imprimir esta página

Xylem SI Analytics AVS 470 Manual De Instrucciones página 47

Equipo medidor de viscosidad
Ocultar thumbs Ver también para SI Analytics AVS 470:

Publicidad

Idiomas disponibles

Idiomas disponibles

Troubleshooting
2.3.8
Check whether the AVS
Trouble
Air bubbles in viscometer
Excessive pumping of measuring
medium into the thermostat bath
or into the safety bottle
®
470 is switch on.
Troubleshooting
 Is the filling quantity sufficient?
- Check, fill viscometer if required.
 Is the viscometer of properly connected?
- in the case of pressing operation, please checked whether the filling
tube is connected; if necessary connect properly.
- for operation in suction mode, please check whether the capillary tube
is connected; if necessary, connect properly.
- please check whether the venting port is tightly connected; if necessary
re-tighten its screwed connection.
 Is the viscometer properly connected?
- for pressing operation?
- for operation in suction mode?
When using AVS
 check position of the rack within the stand
 check the electrical connection from the viscometer to the
ViscoPump type III module
 Is the green LED on the measuring stand illuminated?
When using TC viskometers
 Is the viscometer properly connected?
®
measuring tripods
47

Publicidad

loading