TECHNICAL DATA

6.1

Specifications

Ambient conditions
Indoor use only
Altitude: Up to 2000 m
Specified Temperature range: 5° C to 40° C
Humidity: maximum relative humidity 80 % for temperatures up to 30° C decreasing linearly to
50% relative humidity at 40° C
Mains supply voltage fluctuations: up to ±10% of the nominal voltage
Installation category II
Pollution degree: 2
Operability is assured at ambient temperatures between 5° C to 40° C
Model
Measuring range
Resolution
Error limits/Accuracy
Calibration
Memory
Power supply
Size/weight
Display
Input
Temperature-compensation
Housing
STARTER 3100C Bench Conductivity Meter
ST3100C
0.0 μS/cm...199.9 mS/cm
0.1 mg/l...199.9 g/l (TDS)
0.00 ... 100.0 psu (Salinity)
0 °C...100 °C
Automatic range
0.1 ° C
± 0.5 % of the measured value
± 0.3 ° C
1 point
4 predefined standards
99 measurements
The last calibration data
AC Adapter Input: 100-240V ~ X.XA 50/60 Hz
AC Adapter Output: 12V  X.XA
220 W x 175 D x 78 H mm / 0.75 kg
White backlit Liquid crystal
Mini-Din
Linear: 0.00 %/°C...10.00 %/°C
Reference temperature: 20 & 25 ° C
ABS