Assembly Of The Morcellator Set With Protective Tube (Option 1) - Nouvag TCM3000BL Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 165
5.2.1

Assembly of the Morcellator set with protective tube (option 1)

There is a risk of injury to tissue by the sharp edge at the distal end of the cutting tube.
• Handle the cutting tube carefully.
• If the cutting tube is not used, rotate the protective tube to the NO CUT position.
• If the cutting tube is not used, it must be placed on a flat, sterile surface so it cannot fall and cause
injury
• Check the cutting tube.
• Make sure that the cutting edge of the blade is sharp and undamaged (e.g. no cracks, deformation or
burrs).
• Use only cutting tubes in perfect condition.
1.
Inserting the cutting tube through the gear unit
Remove the seal unit from the gear unit.
Insert the cutting tube through the gear unit. The two
guide pins on the cutting tube must fit into the slot on
the flange of the gear unit.
2.
Fixing the cutting tube to the gear unit
Hold the cutting tube with one hand.
Connect the sealing unit to the gear unit with the bay-
onet closure with the other hand.
3.
Mounting protective tube
Select a suitable protective tube.
Push the protective tube over the distal end of the
cutting tube. The guide pin of the protective tube must
be in the distal slot of the gear unit.
Lock the protective tube in position with a click.
4. Protecting the cutting tube
The distal end of the cutting tube is very sharp.
To prevent injury, the cutting tube must be covered at
all times when it is not in use.
To cover the cutting tube, rotate the protective tube
until the words "NO CUT" are visible in the distal slot
of the gear unit.
The cutting tube is completely covered by the protec-
tive tube.
12