NUK 10.256.345 Operación Manual página 56

Ocultar thumbs Ver también para 10.256.345:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Immunity test
IEC 60601 test level
Surge
1 kV line(s) to line(s)
IEC 61000-4-5
2 kV line(s) to earth
interruptions
<5 % UT (>95 % dip in UT)
and voltage
for 0,5 cycle
variations on
power suppl
40 % UT (60 % dip in UT)
for 5 cycles
input lines
70 % UT (30 % dip in UT)
IEC 61000-4-11
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 sec
Compliance
level
Not
Mains power quality should
applicable
be that of a typical
commercial or hospital
environment.
Not
Mains power quality should
applicable
be that of a typical
commercial or hospital
environment. If the user of the
THixyz series requires
continued operation during
power mains interruptions, it
is recommended that the
THixyz series be powered
from an uninterruptible power
supply or a battery.
- 54 -
Electromagnetic environ-
ment – guidance

Publicidad

Tabla de contenido
loading

Tabla de contenido