5. Open a web browser. In the address field, type https://<transmitter's IP
address>:8443.
• Example: https://172.24.65.229:8443
You can see the IP address of the transmitter on the touchscreen, at the
bottom left corner of the configuration menu
Your web browser may give a certificate error, but you can safely proceed to the
transmitter's IP address. You are next prompted to log in or create users.
Connecting to web interface with a direct
connection between transmitter and computer
• Allen key (4 mm), provided
• Ethernet cable, provided
• A pointed object for pressing the button inside the transmitter.
• Computer with:
• RJ45 Ethernet connector
• Web browser, for example, Firefox, Chrome, or Safari. Using the most
recent version is recommended.
1. Open the 2 hex screws on the cover of the transmitter using a 4-mm Allen key.
2. Open the transmitter cover.
3. Connect the Ethernet cable from the transmitter to the computer.
4. Power up the transmitter.
Powering up takes some minutes.
5. Press the PRESS TO ACTIVATE WEB INTERFACE button with a pointed object to enable
the use of configuration mode IP.
The LED next to the button lights up.
The transmitter's IP address: 192.168.5.20
SERVICE PORT
PRESS TO ACTIVATE
WEB INTERFACE
ETHERNET
.
VS
+
CH1
RS-485
1
PROBE
2
+
24 VOUT
CH2
+
ANALOG
CH3
INPUT
+
CH4
25