3.1
Electrical Specifications
Reference Conditions: 23°C ±5°C, 45 to 80% RH, True RMS for V
A
accuracy are specified from 5% to 100% of range, crest factor 1.4 < CF
AC
< 3 at full scale & CF < 6 at half scale.
AC Amperes
Range
Resolution
40A
400A
1000A
AC Volts
(Auto-Ranging)
Range
Resolution
400V
750V
DC Amperes
Range
Resolution
40A
400A
1000A
DC Volts
(Auto-Ranging)
Range
400V
1000V
8
CHAPTER 3
SPECIFICATIONS
(Auto-Ranging)
0.01A
50/60Hz: 1.9% of Reading ± 5cts
0.1A
60/500Hz: 2.5% of Reading ± 5cts
1A
0.1V
1V
(Auto-Ranging)
0.01A
0.1A
1A
Resolution
0.1V
1V
Accuracy
Accuracy (50/500Hz)
1.5% of Reading ± 5cts
Accuracy (50/500Hz)
2.5% of Reading ± 10cts
Accuracy (50/500Hz)
1.0% Reading ± 2cts
Clamp-on Meter Model 514
and
AC
Overload
1200Arms
Overload
1000Vrms
Overload
1200Arms
Overload
1000Vrms