Operation; Button Functions; Data Hold; Peak Hold Function - AEMC Instruments 514 Manual De Instrucciones

Medidores de tenaza trms
Ocultar thumbs Ver también para 514:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4.1

Button Functions

4.1.1

Data Hold

• The last reading may be held on the display by press-
ing the HOLD button. The
in data hold mode.
• When the held data is no longer needed, release the
hold function by pressing the HOLD button again.
4.1.2

Peak Hold Function

• This meter is built with 1ms peak hold function on A
V
, A
AC
DC
• Calibration is necessary before performing a peak hold

operation.

• To start the calibration process, press and hold the
PEAK button for 2 seconds. "CAL" will display on the
LCD and the offset will be calculated and kept in the
meter.
• After calibration, choose either
Peak Hold button to keep the peak reading.
• Pressing the PEAK button for 2 seconds will return the
meter back to normal operation.
• Once the function range is changed, the meter will need
another calibration for peak measurement.
Clamp-on Meter Model 514
CHAPTER 4
OPERATION
and V
ranges.
DC
symbol will be displayed
or
by pressing the
,
AC
11

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido