Gima THP59J Manual De Uso Y Mantenimiento página 11

Termómetro infrarrojo de oído profesional
Ocultar thumbs Ver también para THP59J:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

11
Cleaning and storage
The probe is the most delicate part of the thermometer. Use with care when cleaning the
lens to avoid damage.
Replace the probe cover after each use to ensure an accurate reading and
avoid cross contamination.
a. Storage temperature range: It should be stored at room temperature between -20~+50°C ,
RH ≤ 85%
b. Transportation temperature shall be less than 70°C, RH ≤ 95%
c. Keep the unit dry and away from any liquids and direct sunlight.
d. The Probe should not be submerge into liquids
If device is accidentally used without probe cover, clean the probe as follows:
a. After the measurement, please use the cotton swab with the Alcohol (70%
concentration) to clean the lens( on the inside of the probe).
b. Allow the probe to fully dry for at least 1 minute.
Holding the thermometer too long may cause a higher ambient
temperature reading of the probe. This could make the body temperature
measurement lower than usual.
Battery replacement
This device is supplied with one lithium cell CR2032x1.
1. Open the battery cover: Insert a pointed object into battery cover pick hole. At the same
time, use thumb to push battery cover out.
2. Hold the device and flip the battery out with a small screwdriver.
3. Insert the new battery under the metal hook on the left side 1 and press the right side 2
of the battery down until you hear a "Click".
4. Replace the battery cover
Keep the battery away from children.
The positive (+) side up and the negative (-) side pointed down
ENGLISH

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

25575

Tabla de contenido