WALLACH WA1000B Instrucciones De Uso página 5

Tabla de contenido

Publicidad

WA1000B™ General Purpose Cryosurgical System (N
INSTRUCTIONS FOR CLEANING LL100
The Wallach LL100 Cryosurgical Freezer can be disinfected by wiping down with commercial disinfectant. This wipe down
procedure can include the black plastic handle and white silicone gas hoses. Other methods involving heat, fluid (cold soak-
ing) or steam are not recommended. The Cryofreezer holder must be sterilized after each procedure.
CARE of CONSOLE
Consoles must never be sterilized. Consoles can be wiped down with an appropriate disinfectant. Prevent fluids from
accumulating on the surface, or from spilling into the internal mechanism.
INSTRUCTIONS FOR HIGH LEVEL DISINFECTION of INSULATION PROBE
The white insulation probe which accepts the cryosurgical tip could have mucous membrane contact and should,
therefore, be disinfected "by wrapping with a cloth soaked in a high-level disinfectant to allow the recommended contact
time. After disinfection, the probe should be rinsed with tap water and dried before use. As with other high-level
disinfection procedures, proper cleaning of probes is necessary to ensure the success of the subsequent disinfection."
INSTRUCTIONS FOR CLEANING STERILIZABLE CRYOTIPS
1. Remove disposable shield from cryotip and discard.
2. Plug cryotip before cleaning and sterilizing. (Plug Orifice does not require sterilization).
3. Scrub tip to remove blood and particles.
4. Rinse with water.
5. Cryotips can be steam sterilized by Gravity or Pre-vacuum using the following procedures:
Set-up
Pre-Exposure
(Conditioning)
Exposure
(Steam Injection)
Post-Exposure
(Exhaust / Dry)
6. Attach new disposable plastic shield to the cryotip, ensuring that the tabs on the shields are correctly fitted into the
grooves on the metal tip.
7. Cryotip is now ready for use.
WARNING: DO NOT USE ETO, E-BEAM OR GAMMA STERILIZATION.
1
Rutala, William A., Ph.D., M.P.H., David J. Weber, M.D., M.P.H., and the Healthcare Infection Control Practices Advisory
Committee (HICPAC), CDC Guideline for Disinfection and Sterilization in Healthcare Facilities, pg. 19, 2008.
IMCRY010-DFU • Rev. A • 4/14
TM
Stage / Process
Cycle Profile
Chamber Purge
Sterilization Temperature
Sterilization Time
Dry Time
4
O) • Directions for Use (English)
2
Parameter Set-Points
Gravity
1 minute
o
132
C (270 F)
15 minutes
15 minutes
1
Pre-vacuum
1 minute
o
132
C (270 F)
4 minutes
20 minutes
Wallach
®
Surgical Devices

Publicidad

Tabla de contenido
loading

Productos relacionados para WALLACH WA1000B

Este manual también es adecuado para:

Ll100900506-1-n2o900506-2-n2o

Tabla de contenido