Tabla 19 Parámetros del bus del bloque de transductores (continuación)
Parameter
CAL_VALUE_SAT
CAL_VALUE_CONC
CALIBRATION_TIMER
CAL_SAMPLE_PRD
CAL_SAMPLE_PRD_
STORED_VAL
CAL_PRODUCT
Alert
HOLD
SENSOCHECK
ALARM_LED_MODE
LAST_ERROR
Description
Shows the value of saturation
calibration
Shows the value of
concentration calibration
Sets the calibration timer (time
in which the device should be
calibrated).
Starts the 1st part of Oxy
product calibration.
Shows the stored value of the
first step of Oxy product
calibration
Sets the value for the 2nd part
of Oxy product calibration.
Sets the device to HOLD mode. R/W 1
Enables or disables
Sensocheck.
Sets the LED to HOLD mode.
Shows the last error.
Data
R/W Bytes
Range
type
R
4
float
R
4
float
000–9999h
R/W 4
float
Default: 0000 h =
disable
0 = Nop
R/W 1
uns8
1 = Sample
Default: 0 if step 1 of
R
float
product calibration
was not started
R/W 4
float
0 = Off
uns8
1 = On
0 = Off
R/W 1
uns8
1 = On
0 = Off
R/W 1
uns8
1 = On
0...100
R
2
uns16
Default: 0 = None