Even if the manufacturers' specification is not known, the
capacitance of a measured length of cable (such as 10 feet) can be
used to determine the capacitance per foot; do not use too short a
length such as one foot, because any error becomes magnified in the
total length calculations.
Sometimes, the capacitance of switches, interconnect cables, circuit
board foils, or other parts, affecting stray capacitance can be critical
to circuit design, or must be repeatable from one unit to another.
Series Vs Parallel Measurement (for Inductors)
The series mode displays the more accurate measurement in most
cases. The series equivalent mode is essential for obtaining an
accurate Q reading of low Q inductors. Where ohmic losses are
most significant, the series equivalent mode is preferred. However,
there are cases where the parallel equivalent mode may be more
appropriate. For iron core inductors operating at higher frequencies
where hysteresis and eddy currents become significant,
measurement in the parallel equivalent mode is preferred.
1.4 Accessories
2 AA Size Ni-Mh Rechargeable Batteries
Shorting Bar
AC to DC Adapter
TL885A SMD Test Probe
TL885B 4-Wire Test Clip (Optional)
TL08C Kelvin Clip (Optional)
Carrying Case (Optional)
19
2 pc
1 pc
1 pc
1 pc