The Power Button functions as follows:
Action
If the MolecuLight i:X is OFF, press and
Switches MolecuLight i:X from OFF to ON.
hold the Power Button for 3 seconds
If the MolecuLight i:X is ON, press and
Switches MolecuLight i:X from ON to OFF.
hold the Power Button for 3 seconds
Switches Violet Wavelength LEDs from ON to OFF automatically (to
conserve battery).
If the MolecuLight i:X is ON during
Fluorescence Imaging Mode for greater
than 90 seconds
Mode and then back to Fluorescence Imaging Mode to activate the Violet
Wavelength LEDs.
Switches the MolecuLight i:X from ON to OFF automatically after 30
minutes (timing out to conserve battery).
If the MolecuLight i:X is ON and you
don't touch MolecuLight i:X for 30
minutes
position when powering the MolecuLight i:X ON, you need to move the
Rocker Switch to Standard Imaging Mode and then back to Fluorescence
Imaging Mode to turn on the Violet Wavelength LEDs.
To reset the MolecuLight i:X, switch the device off, then hold the Power
Reset function (to be used when the
Button for 8 seconds. The System Status LED will blink red initially, then
device or any of its functions are not
yellow and then red again twice. This sequence indicates a reset was
functioning as intended)
performed. Remove finger from button once this sequence is observed.
Press the Power Button again for 2-3 seconds to turn on the device.
Table 8: Power Button functions
6.1.1 Status Indicator LEDs
The MolecuLight i:X is designed with four sensors to guide you in capturing optimal images/videos. Each sensor has a
corresponding indicator light (LED) on the MolecuLight i:X.
6.1.1.1 System Status LED
MolecuLight i: X periodically executes a self-diagnosis to ensure it is operating properly. The System Status LED
indicates the overall operational state of the MolecuLight i:X Imaging Device.
Indicator
Solid Green Light
MolecuLight i:X Imaging Device is functioning as expected.
Solid Yellow Light
You can continue imaging, however follow up with MolecuLight as soon as possible.
The device has malfunctioned. Stop using the device immediately and contact
Solid Red Light
MolecuLight to troubleshoot the problem.
Red Light, Yellow Light, and 2
The device is going through a reset (after the Power Button has been pressed).
Red Light flashes
Table 9: System Status LED
When the device is OFF, no System Status LED color is visible. When you power on the device, the
System Status LED blinks green 5 times for 2 seconds, staying solid green on the 5
Note
indicates that the device is performing its self-diagnostic routine.
6.1.1.2 Range Finder LED
TM
MolecuLight i:X
User Manual Revision 1.0
Effect
Note You will need to move the Rocker Switch to Standard Imaging
Note If the Rocker Switch is in the Fluorescence Imaging Mode
Meaning
MolecuLight i:X is equipped with a pulsed laser-based Range Finder sensor which is used to determine the distance
between MolecuLight i:X and the wound. The Range Finder LED indicates to the user when the device is held at the correct
distance for optimal imaging.
LED Indicator
Green Light
Amber Light
Note
The Range Finder LED should be used when in FL-Mode. It is also strongly recommended to be used when in ST-
Mode to obtain images with similar sizes (for presentation or comparison purposes).
The Range Finder LED should always be used when measuring wound area per Section 10.
Caution
Caution
6.1.1.3 Ambient Light Status LED
FL-images and FL-videos need to be captured in a dark environment for optimal fluorescence image quality. The Ambient
Light Status LED indicates if the level of ambient light in the examination room is optimal for Fluorescence Imaging to aid
users in capturing optimal fluorescence images and videos.
Note
For use with Fluorescence Imaging Mode, view the Ambient Light Status LED immediately after turning off room lights and
prior to toggling the Rocker Switch to Fluorescence Imaging Mode.
Indicator
Green Light The ambient light level in the imaging environment is adequate for fluorescence imaging.
Amber Light
th
blink. Blinking
Note
Caution
Caution
Page 17 of 43
MolecuLight i:X
The optimal distance (8-12 cm) between MolecuLight i:X and the wound being imaged has been
reached. See Figure 8.
The distance between MolecuLight i:X and the wound is out of its optimal range (i.e. closer than 8 cm
or further than 12 cm) and/or the Range Finder sensor is unable to determine the device-to-wound
distance. The user can adjust the device-to-wound distance to obtain a Green Light for imaging.
Table 10: Range Finder LED
The Range Finder LED turns green when the distance between Range Finder and the wound is within 8-12
cm.
Capturing a FL-image or FL-video outside the recommended device-to-wound distance may result in
suboptimal fluorescence and/or image quality.
A defect in the Range Finder's function may result in distorted/inaccurate images/videos. See Section
15 Troubleshooting & Support, question 18 on how to verify the Range Finder function.
The Ambient Light Status LED is only ON in Standard Imaging Mode. It turns off in Fluorescence Imaging
Mode.
Meaning
The ambient light level is too high and fluorescence images should not be captured.
The user should reduce the ambient light settings by turning off the room lights, blocking windows, turning
off computer monitors or other means until the Ambient Light Status LED turns green.
Note If you are unable to obtain optimal ambient light for Fluorescence Imaging Mode, MolecuLight
recommends using MolecuLight DarkDrape (used in conjunction with MolecuLight Adapter). For more
information, refer to Section 4.3 or contact MolecuLight.
Table 11: Ambient Light Status LED
The Ambient Light Status LED is not active or ON when the Violet Wavelength LEDs are turned on
for imaging in Fluorescence Imaging Mode. In Fluorescence Imaging Mode, the Ambient Light
Status LED is OFF.
A defect in the Ambient Light Status LED may result in inaccurate images.
Ignoring the Ambient Light Status LED indicator may result in inaccurate images.
TM
User Manual Revision 1.0
Meaning
Page 18 of 43