Technical Data - Lovibond Tintometer MD100 Manual De Instrucciones

Ocultar thumbs Ver también para Tintometer MD100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Technical Data

GB
Instrument
Light source:
Wavelength accuracy
Photometric accuracy*
Photometric resolution
Power supply
Auto-OFF
Display
Storage
Interface
Time
Calibration
Dimensions
Weight
Ambient conditions
Waterproof
CE
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.
48
Technical Data
triple
wavelength, automatic wavelength selection,
direct reading colorimeter
LEDs,
interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
530 nm
= 5 nm
560 nm
= 5 nm
610 nm
= 6 nm
± 1 nm
3% FS (T = 20° C – 25° C)
0.01 A
4 batteries (AAA / LR 03)
lifetime: approx. 17 hours or 5000 tests
automatic switch off
10 minutes after last keypress
backlit LCD (on keypress)
internal ring memory for 16 data sets
IR interface for data transfer
real time clock und date
user and factory calibration
resetting to factory calibration possible
155 x 75 x 35 mm (LxWxH)
approx. 260 g (incl. batteries)
temperature: 5 – 40 °C
rel. humidity: 30 – 90 % (non-condensing)
floating; as defined in IP 68 (1 hour at 0.1 meter)
Certificate for Declaration of CE-Conformity
at www.lovibond.com