Data Transfer Settings - Mettler Toledo SevenGo Duo pro SG78 Instrucciones De Manejo

Ph/orp/ion/conductivity meter
Ocultar thumbs Ver también para SevenGo Duo pro SG78:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

2. Manual storage
If "Manual Storage" is set, Store appears on the display. Press Store to save endpointed readings.
The endpointed reading can only be stored once. When the data is stored, Store disappears from the
mea sure ment screen.
3. Timed interval readings
A reading is stored to memory every time after a certain interval (3 – 9999 s) defined in the menu
has elapsed. When working in the timed-interval reading mode, the interval by can be defined by
enter ing the seconds. The measurement series stops according to the selected endpoint format or
man u al ly by pressing READ. When timed-interval reading is "on", the DL icon
For readings lasting longer than 15 minutes, switch off the auto-shutoff function. The Auto-off override
icon
appears on the display.

5.5 Data transfer settings

1. Interface
Select to transfer the data in the memory to a PC or to a printer. The meter adjusts the baud rate:
1. Printer
Baud rate: 1200
Data bits: 8
Parity: none
Stop bits: 1
Handshake: none
2. PC
Baud rate: 9600
Data bits: 8
Parity: none
Stop bits: 1
Handshake: none
3. LabX direct*
Baud rate: 9600
Data bits: 8
Parity: none
Stop bits: 1
Handshake: none
2. Printout format
Two different printout formats are available: GLP and Short.
* If LabX direct is selected, the printout format is always GLP and English. LabX PC direct software trans-
lates the received data into the selected PC language as defined in the regional and language options.
Examples: pH
pH/ORP/Ion/Conductivity meter SG78
Setup
appears.