Autoclave

3.
Autoclave
The following accessories are autoclave at 134 ,
Handpiece, BLDC motor,
Internal spray nozzle
Tube Clamp,
Tube Holder,
Motor cap for autoclave
1) Clean the remains such as blood or saliva.
2) Lubricate the head of handpiece with spray after didsassembling.
A1 - Do not use the spray on the motor. It may cause the breakage
A2 - Assemble the motor cap for autoclave on the motor.
A3 - Make sure to use the sterilization bag and seal tightly.
A4 - Put the bag in the sterilization machine and operate at 134
minutes.
(ATTENTION) BLDC motor could be damaged in lubrication or sterilization because
of the moisture inflow or else, so please take care of handling.
Autoclaving the handpiece with blood or other remains may cause the breakage of
the handpiece. Therefore, it is necessary to clean the handpiece before autoclave.
Irrigation tube and Y-connector are disposable. Do not use them again.
4.
Accessories
Hanger
Internal spray nozzle
(lever type)
Internal spray nozzle
(button type)
Tube holder
for 4 minutes or 121
Motor stand
Tube clamp
Y-connector
(button type)
Irrigation tube
for 20