Outer dimensions
(mm)
Weight (g)
Manufacturing
date
Operating
Environment
The infrared thermometer has been tested and conforms to the standard
ASTM E1965-98. ASTM laboratory accuracy requirements in the display
range of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is
±0.4°F (±0.2°C). Note that for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E667-86 and
E1112-86 is ±0.2°F (±0.1°C).

Security Class

Type of protection against electric shock: internally powered equipment.
Degree of protection against electric shock:
■ Degree of protection against ingress of water:IP22
■ Safety degree of using in flammable anesthetic gas blending with air,
oxygen or nitrous oxide: Non-AP/APG
■ No application parts of the thermometer prevents defibrillation charge
effect.
■ No application parts of the thermometer output signal.
■ The thermometer is impermanent installed device.
155.9*40.2*49.2mm
Thermometer (with batteries): 90g
see the label
Temperature: 10°C~ 40°C (50°F–104°F)
Humidity: 15%–95% RH, non-condensing
Atmospheric pressure: 86–106 kPa
26