2 Clean the surfaces using a soft cloth lightly
dampened in a mild soap or detergent
cleaning solution.
Apply the solution to the cloth rather than
the surface.
3 Wipe the surfaces with the disinfecting
solution. Follow the manufacturer's
instructions for the disinfecting solution,
including exposure time and temperature
for effective disinfection.
4 Air dry or towel dry with a clean cloth.
Safety
Observe the following warnings and cautions
before using accessories and peripherals with
the ultrasound system and transducers.
See also the battery safety information in the
ultrasound system user guide.
Electrical safety classification
Class II equipment
The PowerPack is classified
as Class II equipment
when connected to the
AC power supply.
Internally powered
The PowerPack is classified
equipment
as internally powered
equipment when not
connected to the power
supply (battery only).
Non AP/APG
The PowerPack is not
suitable for use in the
presence of flammable
anaesthetics.
Electrical safety
WARNING:
To avoid the risk of electrical shock:
• Use only properly grounded
equipment. Shock hazards exist if
the power supply is not properly
grounded. Grounding reliability
can only be achieved when
equipment is connected to a
receptacle marked "Hospital
Only" or "Hospital Grade" or the
equivalent. The grounding wire
must not be removed or
defeated.
• Use only accessories and
peripherals recommended by
FUJIFILM SonoSite, including the
power supply. Use the PowerPack
only with FUJIFILM SonoSite
power supply P09823.
• Do not disassemble or alter the
battery.
• Use only with specified approved
power supplies.
• Do not plug the stand's power
cord into a multiple portable
socket outlet (MPSO) or use an
extension cord.
To avoid the risk of electrical shock
or injury, do not open the battery
enclosure on the PowerPack battery
and housing.
To avoid the risk of electrical shock
and fire hazard, inspect the power
supply, plugs, cables, and power
cords on a regular basis for damage.
PowerPack User Guide