Descargar Imprimir esta página

About The Control Solution - Accu-Chek Aviva Manual Del Propietário

Ocultar thumbs Ver también para Aviva:

Publicidad

Idiomas disponibles

Idiomas disponibles

Control Tests
When to Perform a Control Test
Performing a control test lets you know the meter and test strips are working properly. You should perform a
control test when:
• you open a new test strip box.
• you left the test strip container open.
• you think the test strips are damaged.
• you want to check the meter and test strips.
• the test strips were stored in extreme temperatures, humidity, or both.
• you dropped the meter.
• your test result does not match how you feel.
• you want to check if you are performing the test correctly.

about the Control Solution

• Use only ACCU‑CHEK Aviva control solution.
• Close the control solution bottle tightly after use.
• Write the date you open the control solution bottle on the bottle label. The control solution must be discarded
3 months from the date the control solution bottle was opened (discard date) or on the Use By date on the
bottle label, whichever comes first.
• Do not use control solution that is past the Use By or discard date.
• Refer to the control solution package insert for control solution storage conditions.
• The meter automatically recognizes the difference between the control solution and blood.
• The control results are not displayed in memory.
• The control solution can stain fabric. Remove stains by washing with soap and water.
• Control solution is available for purchase. To order the control solution, talk to your pharmacist or visit
accu‑chek.com to order online.
12

Publicidad

loading