assure that it is used in such an environment.
Electrostatic discharge
(ESD) IEC 61000-4-2
Electric fast transient /
burst IEC 61000-4-4
Surge IEC 61000-4-4
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
+/- 6 kV contact
+/- 15 KV air
+/- 2 kV for power supply
lines
+/- 1 kV - line
+/- 2 kV - line - earth
<5% UT
(>95% dip in UT) for
0.5 cycle
40% UT
(60% dip in UT) for 5
cycles
+/- 6 kV contact
+/- 15 KV air
+/- 2 kV for power supply
lines
+/- 1 kV - line
No prot. earth
<5% UT
(>95% dip in UT) for
0.5 cycle
40% UT
(60% dip in UT) for 5
cycles
Dispenser
Floors should be
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%
Mains power quality
should be that of typical
commercial or dental
environment.
Mains power quality
should be that of typical
commercial or dental
environment.
Mains power quality
should be that of typical
commercial or dental
environment. If the
user of the VisCalor®
Dispenser requires
continued operation
during power mains
interruptions, it is
recommended that the
VisCalor® Dispenser
be powered from an
uninterruptible power
supply or battery
31