EMC Guidance

1. The Blood Pressure Monitor needs special precautions regarding
EMC and needs to be installed and put into service according
to the EMC information provided in the ACCOMPANYING
DOCUMENTS
2. Wireless communications equipment such as wireless home
network devices, mobile phones, cordless telephones and their
base stations, walkie-talkies can affect this equipment and should
be kept at least a distance d = 3,3 m away from the equipment.
(Note. As indicated in Table 6 of IEC 60601-1-2:2007 for ME EQUIPMENT,
a typical cell phone with a maximum output power of 2 W yields d =
3,3 m at an IMMUNITY LEVEL of 3 V/m)
GUANGDONG TRANSTEK MEDICAL ELECTRONICS CO., LTD
Zone A, 5/F., Investment Building, No. 12, Huizhan East Rd., Torch
Development District, Zhongshan, Guangdong, 528437, China
MDSS - Medical Device Safety Service GmbH
EC REP
Schiffgraben 41, 30175 Hannover, Germany