The system will save the data and increase to 100 ºC
automatically. The cursor will disappear as showed in
chart.
When the system temperature reaches 100 ºC, the
temperature of PV is actual temperature and the tem-
perature of "AdjTemp" is 100 ºC with the cursor. The
"Adj" and "*" is flickering.
After 20 min, the actual temperature of thermome-
ter is 98.0 ºC. Press buttons remove to the left or right
to make the cursor rest on the digital you want to
change. Press increasing or decreasing to adjust.
After the finish of adjustment, the screen displays as
chart.
After the finish of adjustment, press button removes
to the right and ADJ at the same time and keep off at
once. After 8s the cursor display and the system
accepts the adjustment. The screen returns to the dis-
play as the chart.
After the adjustment, the temperature displayed is
the same the actual temperature
NOTE: During the adjustment, press the button removes to the right and ADJ contem-
porary to cancel the adjustment. The system keeps the last adjustment¡, so don't press
casually.
5. Exchange of the metal block
Open the transparent lid and pull out the four screws which fix the block to the heating
board with the screwdriver, close the lid, and put out the block from the main engine.
Take another model of block, steadily lays aside on the main engine. Put in the instalment
holes the screws, and fix with the spanner the metal block on the instrument.
The bath doesn't include any block. The following blocks are available:
Code
Capacity
90603021
96 wells of 0.2 mL
90603022
54 x 0.5 mL
90603023
35 x 1.5 mL
90603024
35 x 2.0 mL
Page 20
Instruction manual 50603010
ENGLISH
PV: 40.1 C
Adj *
AdjTemP=
40.1 C
PV: 100.1 C
Adj *
AdjTemP=
100.0 C
PV: 100.1 C
Adj *
AdjTemP=
98.0 C
PV: 100.1 C
Adj *
AdjTemP=
100.0 C
PV: 98.9c
T:1
SV: 40.0
1:00M
PV: 40.0C
T:1
SV: 40.0
1:00M
Version 1 April-08
3.1 Display de la temperatura actual
3.2 Display de la temperatura estable-
cida
3.3 El primer segmento
3.4 Señal de recuerdo de que la máqui-
na esta trabajando
3.5 Display del tiempo establecido
3. ESPECIFICACIONES TÉCNICAS
1. Condiciones normales de trabajo
- Temperatura ambiente: 5 ºC - 30 ºC
- Humedad relativa: < 70%
- Alimentación: AC 220 V, 50 Hz
2. Parámetros básicos y prestaciones
Referencia
Modelo
Rango de temperatura
Temporizador
Precisión temperatura
Resolución pantalla
Tiempo de calentamiento (de 20 ºC a 100 ºC)
Potencia
Fusible
Medidas (mm)
Peso (kg)
El baño no incluye ningún bloque. Los siguientes bloques están disponibles:
Referencia
Capacidad
90603021
96 x 0.2 mL
90603022
54 x 0.5 mL
90603023
35 x 1.5 mL
90603024
35 x 2.0 mL
Revisión 1, Abril-08
Manual de instrucciones 50603010
CASTELLANO
3.1
3.3
3.4
PV: 100.1c
T: 1*
SV: 100.0c
20:30 M
3.5
3.2
50603010
603/10
RT +5 ºC - 100 ºC
1min - 99h 59 min
< 0.3 ºC
0.1 ºC
12 min
120 W
250V 2.5A ?5×20
300 x 212 x 180
2.5
Pág. 5

Publicidad