Troubleshooting - AND UT-201BLE-A Manual De Instrucciones

Ocultar thumbs Ver también para UT-201BLE-A:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 43
Problem
Nothing appears in
the display, even
when the power is
turned on.
No measurement
Normal body
temperature
includes error.
Body temperature
is displayed low.
Body temperature
is displayed high.
Data transmission
error
Note: If the actions described above do not solve the problem, contact the
dealer. Do not attempt to open or repair this device, as any attempt to do
so will make your warranty invalid.

Troubleshooting

Possible Reason
Battery is drained.
Battery terminals are not in
the correct position.
Battery power is low.
If the battery is drained
completely, the mark does
not appear.
Temperature will change at
awaking, in activity, after
eating.
Incorrect sensor position
The device calculates an
direct mode temperature
after five minutes.
Therefore, it includes error.
The paring is not
established.
Battery is not enough.
English 16
Recommended Action
Replace the battery with a new
one.
Reinstall the battery with negative
and positive terminals matching
those indicated on the battery
compartment.
Replace the battery with a new
one.
Measure the temperature under
the same condition.
Check the sensor position.
Measure after several minutes
again or use the direct mode
measurement.
Place the device in proximity of
the receiver. Make a paring.
Replace the battery with a new
one.