Section 6 - Troubleshooting
Section 7 - Miscellaneous
5.6 Changing the Filter
Filters must be changed only by qualified service
personnel. Refer to the TC3000 Service Manual for
instructions on replacing the filter. Under normal
use, replace the HEPA filter inside the Power Unit
every 1000 operating hours or 12 months,
whichever occurs first.

6.0 Troubleshooting

Most Power Unit operating problems are minor
and can be corrected by the operator. The
following are troubleshooting hints. Perform
corrective action in the sequence listed.
6.1 Quilt Will Not Inflate
1.
Make sure the Power Unit is plugged in and
turned on.
2.
Check both ends of hose for proper
connection.
3.
Check hose and quilt inlet for kinks.
4.
Check Quilt for damage. If air is flowing from
hose, try another Quilt. (Small rips or tears in
the Quilt may be temporarily repaired with
adhesive tape.)
5.
Have qualified service personnel check for a
clogged or dirty HEPA filter.
6.2 REMOVE FROM USE Indicator is Lit
This indicates that the heater and blower have
shut down due to the air temperature exceeding
the high temperature limit, or that the filter has
been incorrectly installed or is missing. Remove
the Power Unit from use immediately. Refer
servicing to qualified personnel.
WARNING
Repairs to the Power Unit should be
performed by qualified personnel such as
certified biomedical electronics
technicians or clinical engineers familiar
with repair practices for servicing medical
devices, and in accordance with TC3000
Service Manual.
Damage to the Power Unit or
malfunction may otherwise result.
8
7.0 Miscellaneous
7.1 Quilts
Contact your dealer for information on the model
TC1050, TC2050, TC2052, TC2054, and TC2060
Quilts available for use with the TC3000 Power
Unit.
WARNING
Use only quilts recommended by
manufacturer. Failure to do so may result
in thermal injury.
7.2 Accessory Stand
An optional accessory stand for mounting the
TC3000 Power Unit is available from your local
dealer.
7.3 Customer Training
In-service training is recommended and can be
arranged through your dealer.
7.4 Limited Warranty
The Thermacare TC3000 Power Unit is warranted
free of defects in material and workmanship under
normal use and operation for a period of two
years, under the terms and conditions of the
Gaymar warranty in place at time of purchase.
Consumable items such as filters are excluded.
The full warranty is available from Gaymar upon
request.
7.5 Return Authorization
Contact your local dealer or Gaymar's Technical
Service Department at:
Phone: 1 800 828-7341
(716) 662-2551
Fax:
(716) 662-8795